- Title
- Rapid UPLC - MS/MS method for the determination of ketoprofen in human dermal microdialysis samples
- Creator
- Tettey-Amlalo, Ralph N O
- Creator
- Kanfer, Isadore
- Date Issued
- 2009
- Date
- 2009
- Type
- text
- Type
- Article
- Identifier
- vital:6444
- Identifier
- http://hdl.handle.net/10962/d1006631
- Description
- Dermal microdialysis (DMD) is a technique capable of determining the percutaneous penetration of drugs from topical formulations intended for local and/or regional activity. Typically, the concentrations of drug collected in dialysates are very low, generally in the ng/ml or even pg/ml range. An additional challenge is the very low volume of sample collected at each collection time and which can range from 1 to 30 μl only. Hence the objective was to develop and validate a rapid, accurate, precise, reproducible and highly sensitive LC–MS/MS method for the quantitative analysis of ketoprofen (KET) in dialystes following application of a topical gel product to the skin of human subjects. UPLC–MS/MS was used and KET was separated on an Acquity™ UPLC BEH C18 column (100 mm × 2.1 mm i.d., 1.7 μm) and analysed in negative-ion (NI) electrospray ionisation (ESI) mode. The mobile phase (MP) consisted of acetonitrile:methanol:water (60:20:20, v/v/v) under isocratic conditions at a flow rate of 0.3 ml/min. Samples were extracted using ethyl acetate with ibuprofen (IBU) as internal standard (IS) and the organic solvent was then evaporated to dryness and the residue re-constituted in methanol. 5 μl samples were injected and analysis was performed at ambient temperature 22 ± 0.5 °C. KET and IBU eluted at 1.07 and 1.49 min, respectively. KET and IBU responses were optimised at the transitions 253.00 > 209.00 and 205.00 > 161.00, respectively. Calibration curves were linear over the range 0.5–500 ng/ml with correlation coefficients > 0.999. The accuracy and precision of the method were found to be between 99.97% and 104.67% (R.S.D. < 2%) and the mean recovery of KET from normal saline was 88.03 ± 0.3% (R.S.D. < 2.20%). The LLOQ and LOD values were found to be 0.5 and 0.1 ng/ml respectively whereas the ULOD was set at 500 ng/ml. The method was successfully applied to determine the bioavailability of KET following application of topical KET gel, Fastum® gel, to the skin of human volunteers.
- Format
- 18 pages
- Format
- Language
- English
- Relation
- Tettey-Amlalo, Ralph Nii Okai and Kanfer, I. (2009) Rapid UPLC–MS/MS method for the determination of ketoprofen in human dermal microdialysis samples. Journal of Pharmaceutical and Biomedical Analysis, 50 (4). Available: http://dx.doi.org/10.1016/j.jpba.2008.09.051
- Hits: 1009
- Visitors: 1094
- Downloads: 107
Thumbnail | File | Description | Size | Format | |||
---|---|---|---|---|---|---|---|
View Details Download | SOURCEPDF | 229 KB | Adobe Acrobat PDF | View Details Download |